- SF cube cell analysis technology for WBC.
- Reticulocyte & NRBC flag detection.
- Provides dedicated flag called “ Infected RBC” ( Malaria screening).
- PLT- Optical + fluorescence measurement.
Principles:
SF cube cell analysis technology for WBC 5-part diff, NRBC, RET and PLT-O.
Focusing flow-DC method for RBC and PLT.
Cyanide free hemoglobin measurement.
Parameters:
33 reportable parameters: WBC, Lym%, Mon%, Neu%, Bas%, Eos%, Lym#, Mon#, Neu#,
Eos#, Bas#; RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, RET%, RET#, IRF, LFR,
MFR, HFR, NRBC#, NRBC%; PLT, MPV, PDW, PCT, P-LCR, P-LCC
14 research parameters: HFC#, HFC%, IMG#, IMG%, WBC-R, WBC-D, WBC-B, WBC-N,
RBC-O, PLT-O, PLT-I, PDW-SD, InR#, InR‰
2 histograms for RBC and PLT
3 scattergrams (3D) for DIFF, NRBC and RET
6 scattergrams (2D) for DIFF, BASO, NRBC, RET, RET-EXT, PLT-O
Performance:
Parameters |
Linearity |
Precision |
Carryover |
WBC |
0-500x 109/L |
<= 2.5% (>= 4x 109/L) |
<=1.0% |
RBC |
0-8x 1012/L |
<= 1.5% (>= 3.5x 1012/L) |
<=1.0% |
HGB |
0-250g/L |
<= 1.0% (110x180g/L) |
<=1.0% |
HCT |
0-75% |
<= 1.5% (30%-50%) |
<=1.0% |
PLT |
0-5000x 109/L |
<= 4.0% (>= 100x 109/L) |
<=1.0% |
RET# |
0-0.8x 1012/L |
<= 15% (RBC >= 3x 1012/L; 1% <= RET % <=4%) |
/ |
Sample Volume:
Predilute mode (capillary blood), open vial 40ul.
Manual mode (whole blood), open vial 150ul.
Autoloader mode (whole blood), closed vial 200ul.
Throughput:
Up to 125 samples per hour (CBC+DIFF).
Up to 90 samples per hour (CBC+DIFF+RET).
Loading Capacity:
Up to 100 sample tubes.
Mode:
CBC, CBC+DIFF, CBC+RET, CBC+NRBC, CBC+DIFF+RET, CBC+DIFF+NRBC, CBC+DIFF+RET+NRBC, RET
Data Storage Capacity:
Up to 100,000 results including numeric and graphical information.
Printout:
Various formats and user-defined formats available.
Operating Environment:
Temperature: 15oC ~ 30oC
Humidity: 30~ 85%
Weight <= 125kg